Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: BK Precision 388-HD: CAL ADJ /5500 DATE: 27-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 69 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # BK Precision Model 388-HD Servicing Information, 1989. # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- J F W 1.003 HEAD EQUIPMENT SETUP 1.004 DISP Turn the UUT on. 1.004 DISP Ambient temperature: 21C - 25C. 1.004 DISP Relative humidity: less than 75%. 1.004 DISP Allow 30 minutes warm up for full stablization. 1.005 HEAD DC VOLTAGE 1.006 DISP Connect the 5500A to the UUT as follows: 1.006 DISP [32] 5500A NORMAL HI to UUT V/Ohm/Hz 1.006 DISP [32] 5500A NORMAL LO to UUT COM 1.007 RNG 200mV 1.008 5500 190.0mV S 2W 1.009 DISP Adjust VR1 for an LCD reading of 190.0 +/-1 digit. 1.010 5500 * S 1.011 HEAD AC VOLTAGE 1.012 RNG 200mV AC 1.013 5500 190.0mV 400H SI S 2W 1.014 DISP Adjust VR2 for an LCD reading of 190.0 +/-1 digit. 1.015 5500 * S 1.016 HEAD CAPACITANCE 1.017 DISP Connect the 5500A to the UUT as follows: 1.017 DISP [32] 5500A NORMAL HI and LO to UUT "CX" socket 1.018 RNG 200nF 1.019 5500 190nF S 2W 1.020 DISP Adjust VF4 for an LCD reading of 190nF +/-1 digit. 1.021 5500 * S 1.022 HEAD FREQUENCY 1.023 DISP Connect the 5500A to the UUT as follows: 1.023 DISP [32] 5500A NORMAL HI to UUT V/Ohm/Hz 1.023 DISP [32] 5500A NORMAL LO to UUT COM 1.024 RNG 2kH 1.025 5500 1.900kH 1Vpp SQ S 2W 1.026 DISP Adjust VR3 for an LCD reading of 1.900 +/-1 digit. 1.027 END